PT Round - PTC/MECH/DGHG/001
PT Item Description - Mahr make 2D Digital Height 0-600 mm (This is an onsite PT activity, where participants needs to carry their instrument to Octagon's Laboratory at Pune and perform measurements).
Accuracy of DUC - (1.8 + L/600) L in mm
Parameter - instrumental error
Delivery - Lab code will be delivered within 7 working days
Participation Criteria - The participant laboratory should have CMC coarser than 11.6 μm
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/PRVC/001
PT Item Description - Analog Hydraulic Pressure Indicating Device, Operating Range 0 to 600 bar, L.C. 5 bar, Accuracy ± 0.5% FS
Parameter - Hydraulic Pressure - Range:0 to 600 bar
Participation Criteria - Participant laboratory should have CMC greater than 0.009% rdg
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/BORG/001
PT Item - Bore Gauge with Dial 18-35 mm with dial 0.001 mm
Parameter - Instrumental Error
Participation Criteria - the participation laboratory should have CMC > 1 μm
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/CALI/001
PT Item- Digital Caliper 0-300 mm with L.C.0.01mm
Parameter- Instrumental Error
Participation Criteria- The CMC of participant lab. should be > 11 μm
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/CYMP/001
PT Item Description - Cylindrical Pin 1 mm, 5 mm and 10 mm
Parameter - Diameter at Marked Point
Participation Criteria - The participant laboratory should have CMC coarser than 0.8 μm
Delivery- Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/DURO/001
PT Item - Shore- A & D Digital Hardness Tester - Range 0-100 Shore , Resolution 1 Shore for A and D
Parameter - Shore A and Shore D (0-100 Shore)
Participation Criteria - The participating Laboratory should have Estimated Uncertainties > 0.6 for Shore A & D
Delivery- Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration- Open
Select options
This product has multiple variants. The options may be chosen on the product page
PT Round - PTC/MECH/MICR/001
PT Item Description - Digital Micrometer - External 0-25 mm
Resolution- 0.001 mm
Parameter - Instrument Error
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Participation Criteria - The participant laboratory should have CMC coarser than 1.1 μm
Note: Participation criteria is defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. However as per ISO/IEC 17043:2010 management system, a standard for PT provider’s accreditation, do not restrict PT provider to include any participant whose CMC’s are superior than reference lab. Octagon provides such choice to the participants to decide their participation in the above PT scheme.
Registration - Open
PT Round - PTC/MECH/TORQ/001
PT Item - Torque Wrench 500 Nm - Measurement Points 50,300,500 Nm
Resolution - 0.1 Nm
Parameter - Torque
Participation Criteria - The CMC of participating lab. should be >0.1%
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Last date of Registration - Open
PT Round - PTC/MECH/FEEG/001
PT Item Description - Feeler Gauge of three different sizes up to 1 mm
Parameter - Thickness
Participation Criteria - The participant laboratory should have CMC coarser than 0.7 μm
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/GABL/001
PT Item - Gauge Block (Tungsten Carbide)
Parameter - Central Length Deviation for 1.5,40and 90 mm
Participation Criteria - The participant laboratory's CMC should be greater than 0.04 + L/ 3000 μm
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/PRVC/004
PT Item Description - Digital Hydraulic Pressure Indicating Device, Operating Range 0 to 700 bar, L.C. 0.01 bar, Accuracy ± 0.1% FS
Parameter - Hydraulic Pressure - Range:0 to 700 bar
Participation Criteria - Participant laboratory should have CMC greater than 0.009 % rdg
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open
PT Round - PTC/MECH/PRVC/005
PT Item Description - Digital Hydraulic Pressure Indicating Device, Operating Range 0 to 200 bar, L.C. 0.01 bar, Accuracy ± 0.1% FS
Parameter - Hydraulic Pressure - Range:0 to 200 bar
Participation Criteria - Participant laboratory should have CMC greater than 0.009 % rdg
Delivery - Lab code will be delivered within 7 working days
Note: Participation criteria defined above, is based on present NABL practice of ILC, where CMC of reference lab should be superior (smaller measurement uncertainty) than the participant’s CMC. All Participant labs should ensure that the participation criteria have been met while applying for the PT. The participant lab should not report estimated uncertainty, superior than accredited CMCs for that parameter for PT measurements. If Estimated uncertainty is coarser to accredited CMCs then Estimated uncertainty should be reported by the participant laboratory.
Registration - Open